Audiometer Specifications

Technical specifications for the Sound Scouts Automatic Audiometer, including test protocols and RETSPL values for supported headphones.

Test Protocols

ParameterValue
Testing protocolScreening (500, 1000, 2000, 4000 Hz) at 20 dBHL; Custom protocol (frequency and levels)
Threshold seeking methodModified Hughson Westlake
Attenuation5 dB step size
Tone type4 burst pure tone
Response window after tone100 ms after tone onset, and 2000 ms after tone offset
Pre-tone waiting periodVariable: 2500 ms–5000 ms
Tone duration235 ms
Rise/fall timeRise/fall times of 35 ms, separated by 235 ms

Reference Equivalent Threshold Sound Pressure Level (RETSPL)

Values for headphones calibrated on an IEC 60318-1 artificial ear. Application force should be that achieved by separating the left and right headphones by 145 mm.

Frequency (Hz)Sennheiser HD 300Sennheiser HD 400STelephonics TDH39RadioEar DD65v2
12530.530.54530.6
25013.713.725.516.9
5006.72.411.57.8
10000.4−7.375.3
20001.4−0.792.3
3000−3.7−3.7102
4000−1.119.59.3
60003.83.115.520.8
80001918.51320.9